Geochemistry

The importance of understanding the geochemical evolution of groundwaters as a basis for isotope hydrogeology cannot be overstated. Interpretations of stable and radioactive isotopes in hydrogeological systems must be complimented by an evaluation of the inorganic and organic geochemistry of groundwater. A variety of special sampling and preservation techniques are required for the analysis of major and certain trace species in natural waters. The following discussion outlines some methods and precautions.

A general practice in sampling water for chemical analysis is to filter the sample through a 0.45-micron pore diameter filter. This will remove most bacteria, almost all suspended clays and a proportion of iron and manganese oxyhydroxides, but will not retain viruses or some organic molecules such as fulvic and humic acids. Filtering is required to assure that the laboratory analysis represents dissolved species (which take part in most geochemical reactions and are used in chemical equilibrium equations) and not suspended constituents, which may be contributed from the wells.

 

Field measurements

Geochemical studies are based on the measurement of the inorganic constituents or species in a groundwater and a series of parameters that control the interactions of these species. Certain parameters such as temperature and pH are difficult to preserve during storage and should be measured in the field.

Temperature: The temperature of groundwater is a fundamental measurement and is required in all thermodynamic calculations and models. It is also needed to correct EC measurements and for calibrating the pH meter. It must be measured as close to in situ conditions as possible. Use a standard glass thermometer, or the temperature probe that is a component of most pH meters.

Electrical conductivity (EC): Electrical conductivity is proportional to the quantity of dissolved ions present in solution and can provide a rough idea of the total dissolved solids (TDS). For most groundwaters, the EC value, in mS/cm corrected to 25oC, is about 50% greater than the TDS expressed as mg/L, and can be estimated according to:

TDS = A ´ EC (mS/cm)

A @ 0.55 in bicarbonate waters

0.75 in high sulphate waters

0.9 in high chloride waters

pH: A reliable pH measurement is one of the most important field parameters to be measured, and must be made with care and patience. pH is an expression of the negative log of H+ activity (pH = –log [H+]). It is fundamental to thermodynamic calculations, and to the interpretation of carbonate and d13C data. In natural waters, pH is generally between about 6.5 and 8. Measurement at a sampling site requires the calibration of the meter with reference buffer solutions of known pH according to the following recommended protocol.

REDOX potential: Redox measurements provide a measure of the electromotive force of a water, or the relative dominance of oxidized vs. reduced species in solution. Electromotive potential (E) is expressed in volts, relative to the standard hydrogen electrode (Eh), which by this convention has zero potential (Eh = 0 volts). It is generally measured with a silver-silver chloride electrode, which has its own standard potential, ER.

As electrodes are simply measuring an electromotive force, there is no calibration. The electrode can, however, be checked against a solution of fixed Eh to determine whether it is functioning. Two solutions with fixed Eh are Zobell’s (ferric-ferrous cyanide solution) and quinhydrone [pH 4 and 7 buffers saturated with quinhydrone powder (~0.2 g/100 mL]:

Zobell’s Solution (Eh = 0.430 V @ 25°C) Quinhydrone solution

0.0033 M K+-Fe2+-CN- solution pH 7 solution = 86 ± 20 mV

0.0033 M K+-Fe3+-CN- solution pH 4 solution = 263 ± 20 mV

M K+/Cl- solution D3 pH (difference between the two solutions) = 177 ± 4 mV

 

Alkalinity: HCO3– and CO3 concentrations are measured by an alkalinity titration. In most natural waters, total alkalinity can be expressed as bicarbonate concentration. CO3 only becomes an important component of DIC above pH 8.5. Other bases can act as proton acceptors, and so alkalinity is defined as:

Alkalinity = mHCO3– + mCO3 + mH3SiO4– + mH2BO3– + mHS– + mOH–

A variety of organic compounds and colloids can also contribute to alkalinity, although HCO3– remains the major contributor in most natural waters. It is determined by titrating with an acid (generally H2SO4) to an end point near pH 4.3. Carbonic acid is a weak acid, and so the titration end point increases to about pH 5 in low alkalinity waters. The volume of acid (and normality N) required is proportional to the alkalinity, which is most often expressed as equivalent concentration of HCO3–:

 

Alkalinity (mg-HCO3–/L) =

Field measurements are greatly aided by using field kits (e.g. Hach, Merck or others) that replace pipettes, burettes and bottles of acid with a microtitrator and syringes of acid. The titration end-point can be identified with a colour indicator (usually Bromcresol), but it is better to use a pH meter and record about 5 to 10 readings below pH 6 (Fig. Chapter 10 -1).

 

 

It is recommended to perform an alkalinity titration in the field, although many will argue that alkalinity is conservative, even if degassing or calcite precipitation takes place. Nonetheless, reactions such as oxidation of Fe2+, degassing of H2S or equilibration with atmospheric CO2 can have an effect. Field titrations with commercially available kits are simple to do and rival laboratory bench auto-titrators in accuracy.

 


SYNOPSIS OF FIELD MEASUREMENTS:

SAMPLE: Ideally, field measurements are made in a flow-through cell with a low, constant flow of groundwater. Measurements can also be made in aliquots taken from a well or spring, although one must consider the effects of atmospheric contact (CO2 degassing, oxidation with O2, etc.).

EQUIPMENT: ·   Flow-through cell, beaker or sample container.

·   pH-mV meter (with or without temperature electrode).

·   pH, redox (ORP) electrodes and spare filling solutions.

·   pH buffers, thermometer.

·   Eh standard solution.

·   EC meter and probe.

·   Alkalinity test kit, sulphuric acid cartridges

·   Filter system (positive pressure) with 0.45-mm filter disks.

·   Deionized water.

pH: ·   Select two buffers that bracket the anticipated pH of the groundwater, i.e. pH 4 and 7, or 7 and 10.

T°C 0 5 10 15 20 25 30
Buffer pH 4 4.00 4.00 4.00 4.00 4.00 4.01 4.02
Buffer pH 7 7.12 7.09 7.06 7.04 7.02 7.00 6.99
Buffer pH 10 10.20 10.18 10.15 10.12 10.05 10.00 9.96

·   Rinse the electrode and shake gently or touch the electrode with an absorbent paper; do not wipe.

·   During a pH measurement, move the electrode slowly to allow it to "see" a different portion of the water at all times. Still water or rapid flow causes drift due to voltage potentials.

·   pH measurements made in poorly buffered solutions (low TDS, low bicarbonate) may show considerable drift, taking 5 to 10 minutes for a stable reading. Instability indicates that the electrode has aged and should be reconditioned or discarded.

·   Store electrode in 2 M KCl solution or in pH 7 buffer solution (short term), or dry (long term).

 

Eh: ·   Test electrode and meter for functioning against standard solution.

    T°C

    0

    5

    10

    15

    25

    35

    50

    ER

    220

    216

    213

    209

    202

    195

    184

ALKALINITY:

·   As little as 5 to 10 mL can be analysed by most equipment. Micro-analytical methods are available in some labs. Standard sample size is 100 mL.

·   Filter sample with 0.45-mm paper, using positive pressure to avoid degassing of CO2. Avoid contact with air with high pH waters.

·   Titrate to an endpoint of pH 4.3, and record acid normality, amount used, and sample volume.

 

Sulphate should be measured if samples are collected for either sulphate or carbonate isotope analyses, to ensure that sufficient sample is collected, and to calculate the amount of barium chloride needed to fully precipitate the sulphate and carbonate. For this purpose a field kit can be used. Alternatively, a simple test for sulphate is to add barium chloride to an aliquot of sample water (adjust pH to below 6) which will turn cloudy if SO42–- is present. Pour slowly into a graduated cylinder (100 mL will do) with a black X on the bottom until the X is no longer visible. The sulphate concentration is roughly determined from the height of water in the cylinder (in cm), using the equation:

Concentration SO4 = –130· log(cm) + 190 mg/L

 


Major anions (Cl–, F–, SO4–, NO3–, Br–)

Analyses of most anions is normally are done in a laboratory by liquid chromatography, although field kits and ion-specific electrodes exist for most anions of interest. HCO3– and CO3 are determined by titration. Less than 5 mL of sample can be analysed for most major anions, although using 25- to 50-mL samples allows repeat measurements, and dilutions for improving precision of some species. Use thoroughly rinsed HDPE or PP bottles. Samples should be filtered (0.45 mm), stored cool, and — obviously — unacidified.

Sulphate analyses on water that contains reduced sulphur species (H2S, HS– or S) can be very erroneous since they oxidize very fast and sulphate readings are then too high. It is mandatory to remove dissolved sulphide, preferably as CdS since its yellow colour is very indicative of the presence of reduced S-species. Filter before analysing for sulphate. A second 25-mL sample is required for total S (H2S and SO4), filtered out of contact with atmosphere. Preserve with 0.25 mL of 30% H2O2 to oxidize all sulphide to sulphate. This measurement for SO4 , less that of the first sample, gives total sulphide.

 

Major, minor and trace metals

All geochemical analyses require measurement of major metals or cations (Ca2+, Mg2+, Na+, K+ and SiO2). Minor (Fe, Mn, Sr and Ba) and trace metals may be required as well. Cations are best analysed in the laboratory by routine methods, including ICP-AES (inductively coupled plasma-atomic emission spectrometry), ICP-MS (ICP-mass spectrometry), AA (atomic absorption) or FE (flame emission), although several field methods are also available. Only a few millilitres of sample are required for ICP instruments that measure multiple wavelengths or masses simultaneously. A normal sample size is 25 to 50 mL.

Carbonates and oxides can precipitate after sampling, and so samples are acidified to keep metals in solution. To ensure that only dissolved species are analysed, the solution must be filtered first. Ultrapure nitric acid is used to acidify to below pH 2. Reagent grade is adequate for major and minor cations only. HCl should be used for brines as it releases hydrogen ions faster. The water must be filtered and should be stored in fresh HDPE or PP bottles. Used bottles are fine too, but should be acid rinsed.

Dissolved organic carbon (DOC)

Sampling for DOC is recommended for any study where redox processes and the carbonate system are important. It is measured in the laboratory by oxidation of an acidified sample (initially degassed of CO2) and measuring the evolved CO2. The sample must be filtered to 0.45 mm at the point of sampling, and acidified to stop bacterial activity. Acidify with HCl, because HNO3 can oxidize organics in the sample prior to analysis. Collect a 10- to 20-mL sample in a glass vial with a non-organic cap seal (aluminum foil, for example). Scintillation vials are ideal and cheap.

 


SYNOPSIS OF GEOCHEMICAL SAMPLING:

SAMPLE VOLUME: ·   25 to 50 mL for anions.

·   25 to 50 mL for cations.

EQUIPMENT: ·   HDPE or PP bottles.

·   Filter system (positive pressure) with 0.45 mm filter disks.

·   Reagent-grade HNO3 acid (or ultrapure for trace metals).

·   Acid delivery system: micropipette or dropper bottle.

·   10% HCl for DOC samples.

METHOD: ·   Take the required field measurements of T, pH etc.

·   Filter sample water into sample bottles, rinsing bottle and cap at least twice with filtered sample water.

·   Fill anion bottle completely (labeled with sample code and as filtered) and cap.

·   Rinse and fill cation bottle (labeled with sample code and as filtered, acidified) leaving 2-mL head space.

·   Acidify with ~5 drops (120 mL) 5 N HNO3 acid for a 50-mL sample, and cap.

·   If collecting a DOC sample, rinse and fill the 20-mL glass vial with filtered sample water, leaving space for addition of 5 drops HCl.